Operating Conditions
Ambient temperature: 0℃~+45℃
Relative humidity: ≤85%RH
Measurement Range and Constant Current Value (RMS)
Resistance: 0~2Ω (10mA), 2~20Ω(10mA), 20~200Ω(1mA)
Voltage: AC 0~20V
Measurement Accuracy and Resolution
Accuracy:
0~0.2Ω≤±3%±1d
0.2Ω~200Ω≤±1.5%±1d
1~20V≤±3%±1d
Resolution:
0.001Ω, 0.01Ω, 0.1Ω, 0.01V
Measurement Error Caused by Auxiliary Ground Resistance and Ground Voltage
Allowable auxiliary ground resistance RC (between C1 and C2)
0~2Ω, 2~20Ω ≤1KΩ
20~200Ω ≤2KΩ
RP (between P1 and P2) <40KΩ Error ≤±5%
Allowable ground voltage (power frequency RMS) ≤5V Error ≤±5%
Power Supply and Consumption
Maximum power consumption ≤2W
DC: 8×1.5V (AA, R6) batteries
AC: 220V/50Hz
Dimensions and Weight
Dimensions: 220mm×200mm×105mm
Weight: ≤1.4kg