JYL Loop Resistance Tester (200A)
Instrument Performance Features: 1. Large test current, fully compliant with national standards for testing contact resistance. 2. High output voltage, wide measurement range, up to 100mΩ.
JYL Loop Resistance Tester (200A) Features:
1. Large test current, fully compliant with national standards for testing contact resistance.
2. High output voltage, wide measurement range, up to 100mΩ.
3. Test current comes from a high-precision 200A switching constant current power supply, no manual adjustment needed, fast and accurate testing.
4. Four-terminal connection method effectively eliminates the influence of test lead resistance on test results.
5. Fast test speed, displays data in about 5 seconds.
6. Long-term continuous operation, equipped with overheat protection circuit, reliable and stable.
7. Easy operation, lightweight, suitable for field work.
JYL Loop Resistance Tester (200A) Technical Parameters:
1. Test current: 50A, 100A, 200A
2. Range: at 50A 0~100 mΩ
at 100A 0~ 20 mΩ
at 200A 0~ 4mΩ
3. Minimum resolution: 0.1μΩ
4. Accuracy: 0.5%±0.2µΩ
5. Output voltage: DC10V
6. Working mode: continuous
7. Power supply: AC220V±10% 50Hz±1Hz
8. Power consumption: ≤1000W
9. Operating temperature: -20℃~40℃
10. Relative humidity: ≤80%RH, non-condensing
11. Weight: 8.7kg
12. Dimensions: 350mm×230mm×260mm
Display digits | 4 1/2 digits | ||||
Configuration | 200A | 100A | |||
Output current | 200A | 100A | 50A | 100A | 50A |
Range | 0~4mΩ | 0~20mΩ | 0~100mΩ | 0~20mΩ | 0~100mΩ |
Accuracy | 0.5%±0.2μΩ | ||||
Minimum resolution | 0.1μΩ | ||||
Output voltage | DC10V | ||||
Working power supply | AC220V±10%,50Hz±1Hz | ||||
Operating temperature | -20℃~40℃ | ||||
Relative humidity | ≤80%RH, non-condensing | ||||
Dimensions | Length 350mm Width 230mm Height 260mm | ||||