Product Introduction of Zinc Oxide Arrester Resistive Current Tester

   The Zinc Oxide Arrester Characteristic Tester is used for the measurement and analysis of leakage current of zinc oxide [MOA]. It is mainly used to measure the resistive current, thereby analyzing the degree of aging and moisture absorption of zinc oxide. On-site live testing complies with the requirements of the People's Republic of China electric power industry standard "DL474.5—92 Guidelines for Implementation of On-site Insulation Tests — Arrester Tests". This Zinc Oxide Arrester Characteristic Tester can also be used in the laboratory for factory and acceptance tests.


Alternative Names of Zinc Oxide Arrester Resistive Current Tester

Zinc Oxide Arrester Tester, Intelligent Arrester Characteristic Tester, Anti-interference Zinc Oxide Arrester Characteristic Tester, Arrester Resistive Leakage Current Detector, Zinc Oxide Arrester Live Tester, AC Gapless Zinc Oxide Arrester Tester, Zinc Oxide Arrester Characteristic Tester


Product Features of Zinc Oxide Arrester Resistive Current Tester

1. Large-screen LCD display, full Chinese menu operation 

    2. High-precision sampling and processing circuits, advanced Fourier harmonic analysis technology, ensuring data reliability

    3. The reference voltage input end has a 0.1A fuse, and there is also an isolation transformer inside the instrument to completely isolate the instrument from the on-site PT, with double insurance to ensure PT safety

    4. Has resistive current fundamental peak output, side-phase correction and other functions

    5. The Zinc Oxide Arrester Characteristic Tester is equipped with a rechargeable battery, calendar clock, micro printer, and can store 210 sets of measurement data

    6. Equipped with RS-232 communication interface, data can be uploaded to computer for processing


Product Parameters of Zinc Oxide Arrester Resistive Current Tester Varistor Measurement























Technical Indicators



Measurement Range



Maximum Error



Test Conditions



Initial Action Voltage (U1mA)



(0~1700)V



≤±2%±1d



1mA±5μA



Leakage Current (I0.75U 1mA)



(0~199.9)μA



±(2μA ±1d)



0.75 U1mA 

≤±2%±1d>



 Discharge Tube Measurement

















Technical Indicators



Measurement Range



Maximum Error



Test Conditions



DC Breakdown Voltage (Vsdc)



(20~1700)V



≤±2%±1d



Voltage Rise Rate

100V/S±10%



Other Indicators





































Insulation Resistance



6MΩ(500V)



Withstand Voltage



AC1.5kV 50HZ 1min



Operating Temperature and Humidity



0~+40℃,≤85%RH



Storage Temperature and Humidity



-10℃~+50℃,≤90%RH



Power Supply



DC12V Dedicated Power Supply (Core wire + pole)



Power Consumption



8W



Instrument Dimensions



208×190×78mm3



Weight



≤1kg